Journal of the Korean Society of Plastic and Reconstructive Surgeons 2008;35(4):423-430.
Published online July 1, 2008.
CT Number Measurement of Residual Foreign Bodies in Face.
Syeo Young Wee, Hwan Jun Choi, Mi Sun Kim, Chang Yong Choi
Department of Plastic and Reconstructive Surgery, College of Medicine, Soon Chun Hyang University, Kyunsangbuk-do, Korea. medi619@hanmail.net
Abstract
PURPOSE
Computed tomography theoretically should improve detection of foreign bodies and provide more information of adjacent soft tissues. And the CT scanner and PACS program proved to be an excellent instrument for detection and localization of most facial foreign bodies above certain minimum levels of detectability. The severity of injury in penetrating trauma to the face, it is often underestimated by physical examination. Diagnosis of a retained foreign object is always critical. METHODS: From March, 2005 to February 2008 a study was done with 200 patients who had facial trauma. Axial and coronal CT images were obtained with a General Electric(Milwaukee, Wis) 9800 CT scanner at 130kV, 90mA, with a 2-mm section thickness and a 512x512 matrix.
RESULTS
Axial and coronal CT images at various window widths should be used as the first imaging modality to detect facial foreign bodies. The attenuation coefficients for the metallic and nonmetallic foreign bodies ranged from -437 to +3071 HU. As a general rule, metallic foreign bodies produced more Hounsfield artifacts than nonmetallic foreign bodies, thus providing a clue to their composition. All of the metallic foreign bodies were represented by a single peak and had a maximum attenuation coefficient of +3071 HU. Of the nonmetallic foreign bodies, glass had an attenuation coefficient that ranged from +105 to +2039, while plastic had a much lower coefficient that ranged from -62 to -35. wood had the lowest range of attenuation coefficients: -491 to -437.
CONCLUSION
The PACS program allows one to distinguish metallic from nonmetallic foreign bodies and to individually identify the specific composition of many nonmetallic foreign bodies. This program does not, however, allow identification of the specific composition of a metallic foreign body. We recommend this type of software program for CT scanning of any patient with an injury to the face in which a foreign body is suspected.
Keywords: CT; Foreign body; Attenuation coefficient; Hounsfield unit
TOOLS
Share :
Facebook Twitter Linked In Google+ Line it
METRICS Graph View
  • 1,160 View
  • 0 Download

120 Cases of Facial Foreign Body Granuloma.2010 May;37(3)



ABOUT
ARTICLE & SPECIALITY
Article category

Browse all articles >

Speciality

Browse all articles >

BROWSE ARTICLES
AUTHOR INFORMATION
Editorial Office
101-2003, Lotte Castle President, 109, Mapodaero, Mapo-gu, Seoul 04146, Korea
E-mail: apsedit@gmail.com                

Copyright © 2024 by Korean Society of Plastic and Reconstructive Surgeons.

Developed in M2PI

Close layer
prev next